Verification and validation of the finite-volume solver

Question. How far can the finite-volume solver’s answers be trusted, and for what? This page organises the evidence along the lines of ASME V&V 20 (verification and validation in CFD) and ASME V&V 40 with the FDA’s credibility guidance (models used for medical decisions). It says what is established, what is prepared but not yet run, and what is missing. Details and numbers are on Finite-volume solver.

Status (30 September 2026). Code verification is done for the CPU reference solver and for the GPU kernels that exist. Validation against benchmarks with published reference data is done for laminar steady flow. Solution verification tools are built, and so are the cross-code and experimental validation cases, but the runs that matter (the coronary tree, the FDA nozzle) wait for the GPU solver.


Context of use

Hemodynamics in image-derived vessels: laminar, incompressible, isothermal blood flow (Newtonian or shear-thinning), steady or pulsatile, with flow splits, pressure drops and wall shear stress as the quantities of interest. The first application is the HiP-CT coronary tree, compared with the collaborator’s Ansys CFX run on the same mesh.

1. Code verification: does the code solve its equations, at its order?

Evidence

Result

Where

Geometry identities (volumes, closed control volumes)

exact to round-off on every mesh, the coronary mesh included

Numerics

Solutions the scheme must reproduce exactly (hydrostatics, Couette)

round-off on every affine element

Finite-volume solver

Manufactured solutions: slip walls, no-slip walls, flow crossing a pressure outlet with prescribed traction, Carreau–Yasuda with the full stress, smoothly warped and unstructured (Delaunay) meshes

velocity second order everywhere; pressure 1.5–2.1 (1.5–1.7 with no-slip walls, the equal-order limit)

Manufactured solutions

Exact Navier–Stokes solutions: Kovasznay (steady), Womersley and Ethier–Steinman (unsteady)

second order in space; BDF1 first and BDF2 second order in time

Exact flows

Invariances: rotation, reflection, translation, renumbering, dynamic similarity, mirror symmetry

round-off (the High Resolution limiter is component-wise: 8 × 10⁻⁴ under rotation)

Invariances

Conservation: mass and momentum balances from consistent reactions

round-off on every run; wall force = Δp × area exactly in Stokes pipe flow

Invariances

Time-step independence of steady states (false time step, marching to steady state)

10⁻¹³

Invariances

GPU kernels against the CPU reference (gradients, momentum diagonal, residual), fuzzed on all element types

3 × 10⁻¹⁶ relative; deterministic

GPU

Lagged Rhie–Chow gradient (the GPU and CFX strategy) against the implicit form

same solution to 10⁻¹², in 58–61 outer iterations

Invariances

Open in code verification. Structured tetrahedral boxes with prescribed sliding velocity on every face carry large boundary-pressure errors, which pollute the velocity (5× when the exact boundary pressure is imposed). No-slip walls and flow-through boundaries do not show it. The GPU linear solve and outer loop do not exist yet, so they are unverified.

2. Solution verification: how much numerical error is in a given answer?

Tool

Status

Iterative convergence: RMS change, mass and momentum balances per iteration

built

Grid convergence index (Roache; Celik et al. 2008), observed order, asymptotic check (zvcfd.verification)

built, tested

Uniform conforming refinement of any mesh (zvcfd.mesh.refine): the route to a grid study of a mesh that cannot be remeshed

built; demonstrated on a wall-wedge / tet-core pipe

Time-step study (halve and double, CFX MG §16.4.2)

built (transient solver); run on Womersley and Ethier–Steinman

The coronary tree: residual targets 10⁻⁴/10⁻⁵/10⁻⁶, a three-level GCI on a cut-out subtree (5.7 M → 45 M nodes by refinement), and a two-level GCI on the whole tree

pending the GPU solver

3. Validation: are these the right equations and inputs?

Case

Reference

Result

Status

Kovasznay flow, Re = 40

exact

second order, 5 × 10⁻⁴ at 32 × 48

done

DFG 2D-1 cylinder, Re = 20

Schäfer & Turek (1996); FeatFlow

c_D +0.23 %, Δp +0.35 %, c_L +1.6 % at 21,680 nodes

done

Lid-driven cavity, Re = 100, 400

Ghia, Ghia & Shin (1982)

centreline RMS deviation 0.003–0.007 at 64²

done

FDA benchmark nozzle, Re_throat = 500

multi-laboratory PIV (Hariharan et al. 2011)

mesh generator built (fda_nozzle), 46 k nodes

pending: the GPU solver, and the PIV data (the FDA’s hub, nciphub.org, no longer resolves)

Stenosis with separation

Ahmed & Giddens (1983)

not started

planned

4. Cross-code comparison on identical meshes

Comparison

Result

Status

OpenFOAM v2506 on the same mesh (pressure-driven pipe, wall wedges and tet core)

OpenFOAM reads zvCFD’s Fluent files (checkMesh: Mesh OK). At Re = 50, against the exact flow rate (developed Poiseuille on the polygon): zvCFD −0.70 % and −0.53 %, OpenFOAM −6.8 % and −5.8 % at nc = 6 and 12. The earlier 10 % gap, zvCFD above the exact rate, was zvCFD’s momentum entering through the pressure inlet, fixed on 2 October

done

Developed pipe through pressure boundaries (exact at every Re)

converges on extruded meshes (+0.56, +0.20, +0.12 % against Stokes at nc = 6, 12, 18) and Delaunay tetrahedra (−0.59, −0.46, −0.23 % at 12, 18, 24); structured tetrahedra leave a 0.5 % floor for every second-order treatment; the opt-in full reconstruction is unstable on coarse meshes (central differencing along the normal)

done; the structured-tetrahedron floor is a property of that mesh

OpenFOAM on the coronary mesh

the OpenFOAM run exists (OpenFOAM)

pending the GPU solver

Ansys CFX (collaborator), same mesh and boundary table

—

pending: the GPU solver, the CFX .out file

svMultiPhysics on VMR 0066, same .vtu

mesh reader built (zvcfd.mesh.vtk)

pending: svMultiPhysics is not installed

zvCFD LBM against zvCFD FV under refinement

both verified against exact pipe flow separately

pending: joint refinement on one geometry

5. What would still be needed for a regulatory-grade credibility case

Beyond the pending runs: uncertainty quantification of the inputs (segmentation, inlet flow, outlet resistances, blood rheology), which V&V 40 weighs against the model’s influence on the decision. And a validation comparator closer to the context of use (an in-vitro coronary phantom or in-vivo flow data). None of that is started.

Reproducing

PYTHONPATH=. python benchmarks/validation/fv_cases.py            # every case (hours on one core each)
PYTHONPATH=. python benchmarks/validation/fv_cases.py kovasznay  # one case
FOAM_SIF=esi2506.sif PYTHONPATH=. python benchmarks/fv/openfoam_same_mesh.py
pytest tests/test_fv_validation.py tests/test_fv_properties.py tests/test_fv_gpu.py

References

ASME V&V 20-2009, Standard for Verification and Validation in Computational Fluid Dynamics and Heat Transfer; ASME V&V 40-2018, Assessing Credibility of Computational Modeling through Verification and Validation: Application to Medical Devices; US FDA, Assessing the Credibility of Computational Modeling and Simulation in Medical Device Submissions (guidance, 2023); P. J. Roache, J. Fluids Eng. 116, 405 (1994); I. B. Celik et al., J. Fluids Eng. 130, 078001 (2008); P. Hariharan et al., J. Biomech. Eng. 133, 041002 (2011); S. F. C. Stewart et al., Cardiovasc. Eng. Technol. 3, 139 (2012); U. Ghia, K. N. Ghia, C. T. Shin, J. Comput. Phys. 48, 387 (1982); S. A. Ahmed, D. P. Giddens, J. Biomech. 16, 505 (1983).